Posts

Showing posts from January, 2019
Image
STANDARD OPERATING PROCEDURES  Safe Autoclave Operations      The purpose of this document is to provide standard operating procedures for the safe use of autoclaves. Autoclaving is a process used to destroy microorganisms and decontaminate biohazardous waste and microbiological equipment used at Biosafety Level 1 / 2. Autoclave Make/Model: EX:  •Consolidated Stills & Sterilizers, Model # SSR3APB Serial #42090 Person Responsible and Phone #: ______________ Located in Building/Room: Building ____, room ____ HAZARDS       Autoclaves use high pressure (15 psi) and high temperature(121 C) or 250 F)  steam for sterilization. Potential safety risks for the operators include:           · Heat burns from hot materials and autoclave chamber walls and door.           · Steam burns from residual steam coming out from autoclave and materials           on completion of cycle.           · Hot fluid scalds from boiling liquids and spillage in autoclave.           ·